Visual or automated dipstick testing for proteinuria in pregnancy?

TitleVisual or automated dipstick testing for proteinuria in pregnancy?
Publication TypeJournal Article
Year of Publication2017
AuthorsCorrea, M, Côté, A-M, De Silva, D, Wang, L, Packianathan, P, von Dadelszen, P, Magee, LA
JournalPregnancy Hypertens
Volume7
Pagination50-53
Date Published2017 Jan
ISSN2210-7797
KeywordsAutoanalysis, Cohort Studies, False Positive Reactions, Female, Humans, Pre-Eclampsia, Pregnancy, Prospective Studies, Proteinuria, Reagent Strips, Sensitivity and Specificity
Abstract

OBJECTIVES: To compare the Multistix 10SG/visual-read with two automated methods (Multistix 10SG/Clinitek 50 and Chemstrip 10A/Urisys 1100) to detect significant proteinuria among high-risk pregnant women.

STUDY DESIGN: Prospective cohort study at British Columbia Women's Hospital & Health Centre, Vancouver, Canada.

MAIN OUTCOME MEASURES: Diagnostic accuracy determined by sensitivity, specificity, positive and negative likelihood ratios (LR+ and LR-).

RESULTS: 303 (89.6%) of 338 women had a urine sample tested by all three dipstick methods. 196 samples (64.7%) were collected in the morning (subsequent to their first void) and from outpatients. 107 samples (35.3%) were from inpatients at various times throughout the day. A PrCr ⩾30mg/mmol was present in 46 (15.2%) samples. The sensitivity for proteinuria was higher with Multistix 10SG/Clinitek 50 (65.2%) than with Multistix 10SG/visual-read (41.3%, p<0.001) or Chemstrip 10A/Urisys 1100 (54.3%, p=0.06). Specificity was >90% for all methods studied, although it was highest for Multistix 10SG/visual-read (98.4%) compared with either Multistix 10SG/Clinitek 50 (92.6%, p<0.001) or Chemstrip 10A/Urisys 1100 (95.7%, p=0.04). For all methods, LR+ was good-excellent (>5), but LR- poor-fair (>0.20). 29 samples were discordant for proteinuria between methods. 28/29 women had negative proteinuria by Multistix 10SG/visual-read, but at least 1+ proteinuria by an automated method; 17/28 were false positives and 11/28 true positives.

CONCLUSIONS: Automated dipstick methods are more sensitive than visual urinalysis for proteinuria, but test performance is still only poor-fair as a 'rule-out' test for proteinuria. Whether the enhanced sensitivity would be worth the false positives, cost, and personnel training remains to be determined for detection of low-level proteinuria in pregnancy.

DOI10.1016/j.preghy.2017.01.005
Alternate JournalPregnancy Hypertens
Citation Key599
PubMed ID28279448